| Single Stage | Double Stage | |
| Points of Use @ 800mL/min consumption flow of the hemodialysis system | Up to 60 Points of Use | Up to 50 Points of Use | 
| Dialysis water outlet capacity at +15 degrees C | 500/L/h to 3000L/h | 500/L/h to 2500L/h | 
| Dimensions in mm | 1500 mm × 550 mm × 951 mm | 1500 mm × 550 mm × 951 mm (first stage) | 
| Footprint in m² | 0.52 m² | 1.04 m² | 
| Weight in kg (Filled) | 160 kg to 360 kg | 330 Kg to 650 kg | 
| Operating output pressure | max. 6 bar | |
| Distribution loop connection | Direct PE-Xa connector 25 × 3.5 (feed and return) | |
| Noise level | Noise level in SUPPLY mode: 62 - 73 dB (A); | |
| Electrical supply / three-phase current | 208 V 60 Hz; 3 / N / PE Hard wired 20 A | 208 V 60 Hz; 3 / N / PE 2X hardwired 20A | 
| Efficiency / Yield / Recovery | Up to 75% adjustable water conversion factor | |